Comparing HemoCue® and Quantitative Buffy Coat® and Coulter Counter-measured haemoglobin concentrations in African children with acute uncomplicated malaria: a Bland-Altman analysis.


ABSTRACT:

Background

Anaemia is a deleterious consequence of malaria, and its accurate diagnosis is crucial for effective management. However, laboratory methods for measuring haemoglobin (Hb) concentration, like the Coulter Counter and the Quantitative Buffy Coat® (QBC®), are costly and not widely accessible in resource-limited settings. The point-of-care HemoCue® test is a cheaper alternative and suitable in rural areas. The study aimed to determine the level of agreement between Coulter Counter/QBC® vs. HemoCue®-measured Hb concentrations by Bland-Altman analysis.

Methods

As part of a randomized, placebo-controlled trial of single low-dose primaquine in Ugandan and Congolese children with acute uncomplicated Plasmodium falciparum malaria, Hb concentrations were measured on days 0,