Does Leak Matter? A Novel Dynamic Leak Model to Simulate Leak for Performance Testing of Manual Neonatal Resuscitation Devices. A Bench Study.


ABSTRACT:

Introduction

Newborn resuscitation is commonly performed in the presence of face mask leak. Leak is highly variable, pressure-dependent and often unrecognized. The effectiveness of resuscitation devices to deliver adequate inflations in the presence of leak is unknown. Bench models simulating continuous leak have the disadvantage of not accurately reflecting leak occurring during clinical resuscitation. A dynamic leak model based on pressure-release valves was thus developed.

Methods

This study investigates self-inflating bag (SIB) and T-piece resuscitator (TPR) ventilation performance in the presence of dynamic (DLM) compared to continuous (CLM) leak models in a bench study. Five predefined leak levels were tested for each leak model (0%-87%). Resuscitation devices were con