Unknown

Dataset Information

Development and analytical performance of a new research use only (RUO) GP73 automated immunoassay.


ABSTRACT: Golgi protein 73 (GP73) is a serum marker with potential applications in the assessment of chronic liver disease progression. We developed a research use only (RUO) GP73 immunoassay to detect serum GP73 concentration. A pair of internal antibodies that most effectively capture and detect GP73 were used to develop a prototype assay. An internal recombinant GP73 standard was used to prepare the assay calibrator and controls. The stability performance of the antibodies and GP73 calibrator were evaluated. We analyzed assay performance on the automated Alinity i system, including precision, sensitivity, linearity, endogenous interference, and HAMA/RF interference. The RUO Alinity i GP73 immunoassay showed good stability when the reagent kit was stored on board the instrument for more than 30 da

SUBMITTER: Ning S 

PROVIDER: S-EPMC11910086 | biostudies-literature | 2025 Mar

REPOSITORIES: biostudies-literature