Exploiting Molecular Ions for Screening Hydrophobic Contaminants in Sediments Using Gas Chromatography-Atmospheric Pressure Chemical Ionization-Ion Mobility-Mass Spectrometry.
Ontology highlight
ABSTRACT: Hydrophobic organic contaminants (HOCs) are conventionally screened by matching electron ionization (EI) mass spectra acquired using gas chromatography-mass spectrometry (GC-MS) to reference spectra. However, extensive in-source fragmentation hampers de novo structure elucidation of novel substances that are absent from EI databases. To address this problem, a new method based on GC-atmospheric pressure chemical ionization (APCI) coupled to ion mobility-high resolution mass spectrometry (IM-HRMS) was developed for simultaneous target, suspect, and nontarget screening of HOCs. Of 102 target chemicals, 85.3% produced (quasi-)molecular ions as base peaks, while 71.6% displayed method detection limits lower than those of GC-EI-low resolution MS. The optimized method was applied to standard ref