ABSTRACT: Background
OLIF (oblique lumbar interbody fusion) is a minimally invasive surgery to treat spinal instability. However, clinical studies indicated the early degeneration of adjacent segments after surgery. The rod stiffness of OLIF was associated with change at adjacent segments. Therefore, the study aimed to compare the biomechanical effects of OLIF with different rod material properties using the finite element (FE) method.Methods
A validated L1-L5 lumbar spine was conducted in the biomechanical analysis using FE software ANSYS. The FE model of OLIF with a rod was created. Current biocompatible materials for the rod of the OLIF model were changed, including titanium alloy (OLIF_Ti), nickel-titanium alloy (OLIF_NiTi), and polycarbonate urethane (OLIF_PCU) rod. Four FE mode