ABSTRACT: Purpose
A fast T1 mapping method of the whole pancreas remains a challenge, due to the complex anatomy of the organ. In addition, a technique for pancreas water T1 mapping is needed, since the T1 is biased in the presence of fat. The purpose of this work is to accelerate the acquisition of water selective T1 (wT1) mapping for the whole pancreas at 3 T.Methods
The proposed method combines a continuous inversion-recovery Look-Locker acquisition with a single-shot gradient echo spiral readout, water-fat separation and dictionary matching for wT1 mapping of the whole pancreas at 3 T. The bias of T1 in the presence of fat was evaluated in a phantom by comparing the modified Look-Locker inversion-recover