ABSTRACT: Background
Reliable dosimetry based on SPECT/CT imaging is essential to achieve personalized 166Ho-radioembolization treatment planning and evaluation. This study quantitatively evaluates multiple acquisition and reconstruction protocols for 166Ho-SPECT imaging based on data from five Dutch hospitals. We aim to recommend an imaging protocol which harmonizes 166Ho-SPECT images for reproducible and accurate dosimetry in a multi-scanner and multi-center setting.Methods
Cylindrical and NEMA IEC phantoms, filled with 166Ho-chloride, were imaged using seven SPECT/CT scanners from two vendors (GE HealthCare and Siemens Healthineers). Data were acquired with a photopeak window centered at 81 keV. Two adjacent scatter windows, and one uppe