A critical step toward far-field laboratory diffraction contrast tomography in Laue focusing geometry.


ABSTRACT: Recent simulations have demonstrated the promising potential of far-field laboratory diffraction contrast tomography (FF-LabDCT) in the Laue focusing geometry for 3D mapping of grain-averaged deviatoric strain tensors, achieving a strain uncertainty as low as 1 × 10-4 [Lindkvist & Zhang (2022). J. Appl. Cryst. 55, 21-32]. The present work takes a critical step towards realizing this tool. It presents the first attempt at implementing FF-LabDCT using a Zeiss Xradia 520 Versa X-ray microscope, including both hardware and software development. A new algorithm, LabDBB, which adapts the previously developed dictionary-based branch and bound (DBB) principles, is implemented for indexing FF-LabDCT data. The performance of LabDBB is evaluated by comparing the indexed grain

SUBMITTER: Zhang Y 

PROVIDER: S-EPMC11957408 | biostudies-literature | 2025 Apr

REPOSITORIES: biostudies-literature