Additional Screw Added to the Femoral Neck System Could Enhance the Stability of Pauwel Type III Femoral Neck Fractures: a Finite Element Analysis.


ABSTRACT:

Backgroud

This study explores effective fixation methods for Pauwel type III femoral neck fractures by evaluating the biomechanical benefits of adding a screw to the Femoral Neck System (FNS).

Methods

Computed tomography (CT) scans of an 82-year-old female patient with an intertrochanteric fracture were used to establish a finite element femur model with heterogeneous material properties. Finite element models of Pauwel type III fractures were created with and without an additional screw. The central and inferior trajectories of the FNS bolt were examined separately and combined with an additional screw for virtual fixation. Walking and stair-climbing loads were applied.

Results

With the addition of a screw, both peak maximum and minimum principal strains consistently