Correction of metal artefacts around orthodontic mini-implants - a micro-CT study in the rat tail model.
Ontology highlight
ABSTRACT: Micro-CT enables volumetric analysis of peri-implant tissue, but grey value alterations due to metal artefacts can impair analyses. This study aimed to assess to which extent peri-implant grey values are affected by metal artefacts at increasing distance to the implant, and whether mathematical correction is possible. In nine rats, two Ti6Al4V orthodontic mini-implants (OMIs), 0.8 mm in diameter and 3.0 mm in length, were placed in a single tail vertebra. Micro-CT scans were performed before (T0) and after (T1) careful removal of the OMIs. Consecutive micro-CT scans were registered and differences in local grey values were computed at increasing distance to the implant (10.4 μm to 405.6 μm). Correction coefficient (CC) computation was performed using a smoothing spline fit, with the distan