ABSTRACT: Background
Dynamic fetal cardiovascular MRI (CMR) enables visualization of moving structures to assess congenital heart disease and plan treatment. Low field MRI systems can provide more comfortable platforms for fetal CMR. Here, we demonstrate the feasibility and utility of motion corrected fetal cardiac cine CMR and compare it with real-time CMR at multiple spatial resolutions at 0.55 T.Methods
Ten human pregnancies were scanned at 0.55T on a derated MAGNETOM Aera (Siemens Healthineers, Erlangen, Germany) with spiral steady-state free precession imaging. Real-time images were reconstructed and used for motion correction and fetal cardiac gating followed by cine reconstructions. The signal-to-noise ratio (SNR), image quality, blood-to-myocardium contrast, and contrast-to-n