Unknown

Dataset Information

Continuous High-Throughput Plasma Separation for Blood Biomarker Sensing Using a Hydrodynamic Microfluidic Device.


ABSTRACT: Continuous, cost-effective, high-throughput with admissible yield and purity of blood plasma separation is widely needed for biomarker detection in the clinic. The existing gold standard technique (centrifugation) and microfluidic technologies fall short of meeting these criteria. In this study, a microfluidic device design is demonstrated based on passive hydrodynamic principles to achieve admissible yield and purity plasma samples. Through computational and experimental assessments, it is shown that side channels with varying lengths are required to improve the plasma extraction rate. The optimized side channels in this device design use the formed cell-free layer regions in the expanded areas to extract plasma consistently and efficiently. These Hydrodynamic Continuous, High-Throughput

SUBMITTER: Abouali H 

PROVIDER: S-EPMC11973946 | biostudies-literature | 2025 Apr

REPOSITORIES: biostudies-literature