ABSTRACT: Background
Transporting ventilated patients safely and without loss of efficacy is a challenge. Portable ventilators are generally used to transport critically ill patients, but their performance is often limited.Objective
This study aimed to compare the in-vitro performance of a modern portable respirator with a modern intensive care respirator for different lung settings.Design
An in-vitro testing of a portable and an intensive care respirator.Settings
Anaesthesia Department at the University Children's Hospital Zurich.Main outcome measures
The portable respirator Hamilton T1 was compared with the established intensive care respirator bellavista1000 (BV) while applying different settings with the ASL 5000 (ASL) device. The ASL can simulate neonatal