Unknown

Dataset Information

In Vitro Assessment of Vascular Injury following Stent Retriever Retraction in Clinically Relevant Endothelialized Silicone Models.


ABSTRACT: Mechanical thrombectomy devices have potential to injure the vessel during treatment of acute ischemic stroke. The goal of the current work was to tailor in vitro endothelialized silicone models for stent retriever assessment and to evaluate endothelial injury following treatment by various stent retriever designs and sizes. Clinically relevant neurovascular geometries were first modeled out of silicone, then sterilized, coated with fibronectin, placed in bioreactors, seeded with human endothelial cells, and cultivated under flow. Several sizes of 2 different commercially available stent retrievers were then deployed in, and retracted through, vessels. Vessels were immediately harvested and stained. Endothelial injury, identified as denudation, was quantified by using ImageJ. Results illus

SUBMITTER: Starr I 

PROVIDER: S-EPMC11979809 | biostudies-literature | 2025 Mar

REPOSITORIES: biostudies-literature