Reducing Metal Artifacts in Clinical Photon Counting Detector Computed Tomography-A Phantom Study of an Exemplary Total Hip Arthroplasty.


ABSTRACT:

Objective

To examine how different photon-counting detector (PCD) CT scanning and reconstruction methods affect the volume of metal artifacts and image quality for a hip prosthesis phantom.

Methods

A titanium and cobalt-chromium-molybdenum-alloy total hip prosthesis phantom was scanned using a clinical PCD-CT with a constant tube potential (140 kV) and Computed-Tomography-Dose- Index (7 mGy). Different scan settings were used: with/without tin-filter (Sn), with/without ultra-high resolution (UHR), both individually and combined, resulting in four modes: Quantumplus (Standard), UHR Quantumplus (HighRes), QuantumSn (Standard-Tin) and UHR QuantumSn (HighRes-Tin). Reconstructions included virtual monoenergetic images (VMI) spanning 40-190 keV and polychromatic images, with/witho