Cryoglobulinemia concurrent with Leishmania infantum infection in a dog and its interference with two automated hematology analyzers.
Ontology highlight
ABSTRACT: A 6-year-old, intact male English Bulldog presented for evaluation of weakness. Physical examination revealed mildly pale mucous membranes and ophthalmologic alterations. A complete blood cell count was performed on the Urit Smart V5 and the Sysmex XN-1000V, and showed several hematologic discrepancies between the analyzers. The Urit reported marked leukocytosis and thrombocytosis, and the Sysmex reported marked thrombocytosis on the impedance and optical RET channel but mild thrombocytopenia on the optical PLT-F channel. On the blood smear examination, there was an abundant proteinaceous material compatible with cryoglobulins. The red blood cell exhibited a notable degree of poikilocytosis. The number of leukocytes was found to be correlated with the Sysmex total white blood cell count, w
SUBMITTER: Martinez-Caro J
PROVIDER: S-EPMC12009788 | biostudies-literature | 2025 Mar
REPOSITORIES: biostudies-literature