Unknown

Dataset Information

How Repetition Rate Impacts Detection Limits of Ion Mobility Spectrometers with Field-Switching Ion Shutters.


ABSTRACT: Ion mobility spectrometers are extremely sensitive analytical instruments, consisting of an ionization region and a drift region, usually separated by an ion shutter. Highest sensitivity is reached with field-switching ion shutters, as ions are accumulated in the ionization region while the ion shutter is closed, which defines the reaction time in the ionization region. This study investigates the effect of repetition rate and reaction time on ion formation and detection limits. The results reveal that increasing the ionization source intensity and reaction time increases the signal-to-noise ratio even though less spectra can be averaged in a given overall measuring time. It is shown that the formation of protonated monomers and proton-bound dimers is considerably slower than the formation

SUBMITTER: Lippmann M 

PROVIDER: S-EPMC12019778 | biostudies-literature | 2025 Apr

REPOSITORIES: biostudies-literature