ABSTRACT: Background
Advanced swept-source optical coherence tomography (SS-OCT) devices are the current gold standard for the measurement of ocular biometric and keratometric parameters, which are essential for the calculation and selection of intraocular lenses (IOLs). This study compares the agreement of two SS-OCT devices, the Eyestar 900 (Haag-Streit, Köniz, Switzerland) and the Anterion (Heidelberg Engineering, Heidelberg, Germany).Materials and methods
All patients undergoing cataract surgery or seeking consultation for corneal abnormalities between January 2024 and May 2024 were eligible for inclusion. Both eyes were included in the final analysis.Results
A sample of 86 eyes from 43 patients was analysed. The mean differences (ES-AN) across all data between the Eyesta