Imaging near titanium total hip arthroplasty at 0.55 T compared with 3 T.


ABSTRACT:

Purpose

To compare 0.55 T and 3 T MRI for imaging patients with titanium total hip arthroplasty (THA). Patients with orthopedic metallic implants often require diagnostic imaging to evaluate adjacent tissues. MRI performance measures, including artifact levels and SNR, vary with field strength.

Methods

Six patients with titanium THA were scanned with similar protocols at 0.55 T and 3 T, including proton density (PD) weighted turbo spin echo (TSE), PD TSE with view-angle tilting (TSE + VAT), PD slice encoding for metal artifact correction (SEMAC), and short tau inversion recovery with SEMAC (STIR-SEMAC). Images from both field strengths were scored by two readers and qualitatively and quantitatively compared.

Results

Diagnostic confidence was significantly higher at 0.