Safety Clearance and Artifact Testing of a Nitinol Breast Biopsy Clip in an Ultra-High Resolution (7 Tesla) Magnetic Resonance Imaging Environment.
Ontology highlight
ABSTRACT: Background The lack of safety clearance of several metallic breast biopsy clips in 7 Tesla (T) poses a significant hurdle to using advanced magnetic resonance imaging (MRI) techniques in clinical management or cancer research. Aims This article assesses the Ultracor Twirl clip for safety and imaging artifacts in a 7T MRI scanner. Setting and Design This study can be categorized as a phantom study. Materials and Methods Tests for magnetic susceptibility (translational attraction and torque), MRI-related heating, and artifacts were conducted based on the American Society for Testing and Materials standards. The magnetic susceptibility tests evaluated the scanner's magnetic force that can cause clip movement and rotation. The heating test was conducted with cus