ABSTRACT: Objectives
The purpose of this study was to evaluate how artefacts caused by titanium and zirconia dental implants affect the bone quality assessment in CBCT images. The effect of scan mode and the use of metal artefact reduction (MAR) algorithm on artefacts suppression were taken in consideration.Methods
Titanium and zirconia dental implants were installed in porcine bone samples and scanned with two CBCT devices with adjustments on scan mode and with the use of MAR. The control group consisted of bone sample without implant and scanned with full-rotation scan mode without MAR. Artefacts extension and bone quality around implants were measured by deviation of grey values and bone histomorphometry measurements (trabecular volume fraction, bone specific surface, trabecular t