CFD investigations of a shape-memory polymer foam-based endovascular embolization device for the treatment of intracranial aneurysms.


ABSTRACT: The hemodynamic and convective heat transfer effects of a patient-specific endovascular therapeutic agent based on shape-memory polymer foam (SMPf) are evaluated using computational fluid dynamics studies for six patient-specific aneurysm geometries. The SMPf device is modeled as a continuous porous medium with full expansion for the flow studies and with various degrees of expansion for the heat transfer studies. The flow simulation parameters were qualitatively validated based on the existing literature. Further, a mesh independence study was conducted to verify an optimal cell size and reduce the computational costs. For convective heat transfer, a worst-case scenario is evaluated where the minimum volumetric flow rate is applied alongside the zero-flux boundary conditions. In the flow