ABSTRACT: Purpose
To evaluate the repeatability and reproducibility of QSM of the liver via single breath-hold chemical shift-encoded MRI at both 1.5 T and 3 T in a multicenter, multivendor study in subjects with iron overload.Methods
This prospective study included four academic medical centers with three different MRI vendors at 1.5 T and 3 T. Subjects with known or suspected liver iron overload underwent multi-echo spoiled gradient-recalled-echo scans at each field strength. A subset received repeatability testing at either 1.5 T or 3 T. Susceptibility and R2*$$ {\mathrm{R}}_2^{\ast } $$ maps were reconstructed from the multi-echo images and analyzed at a single center. QSM-measured susceptibility was compared with R2*$$ {\mathrm{R}}_2^{\ast } $$ and a commercial R2-bas