Comparison of Intraocular Pressure Measurements with Goldmann Applanation Tonometry, iCare, and Tono-Pen in Young Children with Anterior Segment Abnormalities Under General Anesthesia.


ABSTRACT: Background: In young patients with suspected elevated intraocular pressure (IOP), examinations under general anesthesia remain the gold standard. This study aimed to compare the reliability of Goldmann applanation tonometry (Perkins), iCare rebound tonometry, and the Tono-Pen in young children under general anesthesia in a clinical setting. Methods: This retrospective study included patients under six years of age requiring an ophthalmic examination under general anesthesia. IOP measurements were performed using all three devices, and central corneal thickness (CCT) was recorded for each patient. Results: A total of 38 eyes of 19 children (mean age, 1.8 ± 2.1 years) were included. IOP values of all three devices ranged from 5 to 43 mmHg, with a mean CCT of 645.6 ± 135