Unknown

Dataset Information

TD-ESI-MS/MS for High-Throughput Screening of 13 Common Drugs and 4 Etomidate Analogs in Hair: Method Validation and Forensic Applications.


ABSTRACT: This study established a dual analytical workflow integrating thermal desorption-electrospray ionization-tandem mass spectrometry (TD-ESI-MS/MS) for rapid qualitative screening and ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) for confirmatory quantification of 17 psychoactive substances and metabolites across six classes (opioids, amphetamine-type stimulants, cocaine, ketamine-type drugs, cannabinoids, and etomidate analogs) in hair matrices. Validation of the TD-ESI-MS/MS method demonstrated its sensitivity (limits of detection: 0.1-0.2 ng/mg) and precision (<19.3%), with matrix effects controlled to <19.6%. The TD-ESI-MS/MS method achieved an analysis time of 1 min per sample, enabling high-throughput screening with a sensitivity >85.7% and a specificity

SUBMITTER: Li M 

PROVIDER: S-EPMC12116117 | biostudies-literature | 2025 Apr