ABSTRACT: Background
Efforts have been joined to set the parameters for the reliability of glucometers, yet once they are on the market, they are not further tested for the maintenance of accuracy, specificity, or precision.Methods
This comparative analytical study investigated the precision of commonly used glucometers in Saudi Arabia, namely Accu-Chek Instant®, On-Call Sharp®, and ConTour®, as well as the effects of vitamin C, acetaminophen, and maltose on glucose readings. Ten milliliters of blood was drawn in lithium heparin from healthy volunteers (n = 9). Six samples were divided into two groups of three. One group was designed for normal glucose levels. The second group was designed for high glucose levels by adding a dextrose solution. The las