Gamma radiation detector selection for CT scanner.
Ontology highlight
ABSTRACT: Three types of gamma radiation detectors associated with distributed electronics namely, NaI (Tl), HPGe and LaBr3(Ce) are compared primarily focusing on electronic noise and scattering noise. Additionally, detectors of same make, material, size and electronics are also compared. A methodology is proposed to select the most suitable detector for computed tomography (CT) among the available options. Standard deviation parameter is employed to estimate electronic noise without performing CT experiment. Kanpur theorem-1(KT-1) is used to estimate the scattering noise quantitatively after verifying its sensitivity to scattering noise. The impact of scattering noise on CT profiles is evaluated using dice similarity dice coefficient. A good resemblance between KT-1 and dice coefficient