Validating a T1-weighted cine MRI for a 1.5T MR-Linac with temporal resolution appropriate for respiratory motion.


ABSTRACT:

Purpose

High temporal resolution cine magnetic resonance imaging (MRI) for the 1.5T Elekta Unity MR-Linac system currently relies on a balanced contrast sequence for motion monitoring (MM) and tumor tracking. Despite its high signal-to-noise ratio (SNR), a balanced contrast sequence does not always provide the ideal contrast for tumor imaging in all situations. Thus, the investigation of other contrast high temporal resolution cine MRI sequences is needed.

Methods

Experiments were conducted to validate the T1-weighting and SNR on a cine MRI sequence with a frame rate of 4 frames-per-second (fps) and sufficient image quality. A ModusQA Quasar MRI4D motion phantom and Alzheimer's Disease Neuroimaging Initiative (ADNI) phantom were used to confirm adequate motion tracking, cont