Unknown

Dataset Information

Laboratory development and validation of vapor phase PFAS methods for soil gas, sewer gas, and indoor air.


ABSTRACT: There is no standard sampling and analysis method for vapor phase per- and polyfluoroalkyl substances (PFAS) that can be routinely applied to soil gas, sewer/conduit gas, and indoor air samples. We have validated a thermal desorption GC/MS/MS method for the measurement of a set of fluorotelomer alcohols and perfluorooctanesulfonamides collected on multi-bed sorbent tubes. Applications to perfluorocarboxylic acids were also evaluated since there is debate regarding under what circumstances these compounds could be observed moving into gas phase. Perfluorooctanoic acid (PFOA) met Method TO-17 calibration requirements when calibrated using National Institute of Standards and Technology (NIST) traceable standard solutions introduced through the thermal desorption system and using multiple reac

SUBMITTER: Hayes H 

PROVIDER: S-EPMC12180925 | biostudies-literature | 2025 Jan

REPOSITORIES: biostudies-literature