ABSTRACT: Background
Phase contrast (PC) cardiovascular magnetic resonance (CMR) is clinically used to quantify flow. The quantification accuracy is diminished by background phase errors. Image-based background phase correction algorithms are commercially available, but their accuracy is still under evaluation. Here, we validate a recently developed non-linear phase contrast correction (nPCcor) algorithm that includes automatic failure mode classification in a large single-vendor multi-scanner retrospective study.Methods
Three hundred forty-six through-plane PC images at the aortic valve (AAo) and pulmonary artery (PA) were acquired on three different GE HealthCare 1.5T clinical MRI scanners. Each PC scan was repeated on a static phantom, and the static phantom-corrected PC series wa