Diagnostic performance of single-lead electrocardiograms from a smartwatch and a smartring for cardiac arrhythmia detection.


ABSTRACT:

Background

Wearable devices are widely used for atrial fibrillation (AF) detection, yet most validation studies include only sinus rhythm or AF, likely overestimating diagnostic performance.

Objective

This multicenter study assessed the performance of automated AF detection and physician interpretation of single-lead electrocardiograms (SL-ECGs) from the Apple Watch and CART Ring.

Methodology

Participants underwent simultaneous 12-lead ECG and SL-ECGs from Apple Watch and CART Ring. Two cardiologists independently adjudicated all ECGs. Apple Watch and CART Ring classified recordings as "AF," "Not AF," or "Unclassified." Diagnostic performance for automated AF detection was evaluated in "worst-case" (all SL-ECGs) and lenient (excluding unclassified SL-ECGs) scenarios.