ABSTRACT: Background
Continuous glucose monitoring (CGM) has the potential to improve glucose control in critically ill patients, provided that its trend accuracy is reliable. We evaluated the trend accuracy of a subcutaneous CGM system (Dexcom G6) compared with intermittent arterial blood gas (ABG) measurements in intensive care unit (ICU) patients receiving insulin.Methods
We enrolled 40 adult ICU patients receiving insulin and organ-supportive therapies. We assessed trend accuracy using the Rate Error Grid Analysis (R-EGA) and the Diabetes Technology Society Trend Accuracy Matrix (DTS-TAM), overall, across different ABG levels, and over time from CGM initiation.Results
A total of 2701 paired CGM-ABG trends were analyzed, with a median (IQR) time difference between readings