ABSTRACT: Background
Carbon-fiber custom dynamic orthoses are used to improve gait and limb function following lower limb trauma in specialty centers. However, the effects of commercially available orthoses on center of pressure progression and patient perception of orthosis smoothness during walking are poorly understood.Methods
In total, 16 participants with a unilateral lower extremity traumatic injury underwent gait analysis when walking without an orthosis, and while wearing monolithic and modular devices, in a randomized order. Device alignment, stiffness, participant rating of perceived device smoothness, center of pressure velocity, and ankle zero moment crossing were assessed.Findings
The modular device was approximately twice as stiff as the monolithic device. Align