Unknown

Dataset Information

Using real-world data from health systems to evaluate the safety and effectiveness of a catheter to treat ischemic ventricular tachycardia.


ABSTRACT:

Background

The ThermoCool STSF catheter is used for ablation of ischemic ventricular tachycardia (VT) in routine clinical practice, although outcomes have not been studied and the catheter does not have Food and Drug Administration (FDA) approval for this indication. We used real-world health system data to evaluate its safety and effectiveness for this indication.

Methods

Among patients undergoing ischemic VT ablation with the ThermoCool STSF catheter pooled across two health systems (Mercy Health and Mayo Clinic), the primary safety composite outcome of death, thromboembolic events, and procedural complications within 7 days was compared to a performance goal of 15%, which is twice the expected proportion of the primary composite safety outcome based on prior studies. The

SUBMITTER: Dhruva SS 

PROVIDER: S-EPMC12306424 | biostudies-literature | 2023 Nov

REPOSITORIES: biostudies-literature