Unknown

Dataset Information

Intraocular pressure and ocular biomechanical parameters vary between generations of the ocular response analyzer in healthy and ectatic eyes.


ABSTRACT:

Introduction

This study evaluated the agreement between a third-generation (G3) ocular response analyzer (ORA) and a first-generation (G1) ORA, and tested the ability of the keratoconus match index (KMI) to identify keratoconus.

Methods

Healthy participants (n = 149 eyes) and participants with keratoconus (n = 78 eyes) were enrolled for this study. Four measurements were taken bilaterally using the G1 and G3 ORA. Goldmann-correlated intraocular pressure (IOPg), corneal-compensated IOP (IOPcc), corneal hysteresis (CH), waveform score, KMI, and waveform parameters area under the first applanation peak (p1area), area under the second applanation peak (p2area), width of the first applanation peak (w1), width of the second applanation peak (w2), height of the first

SUBMITTER: Yuhas PT 

PROVIDER: S-EPMC12328285 | biostudies-literature | 2025

REPOSITORIES: biostudies-literature