Automated ultrasound system ARTHUR V.2.0 with AI analysis DIANA V.2.0 matches expert rheumatologist in hand joint assessment of rheumatoid arthritis patients.


ABSTRACT:

Objective

To evaluate the agreement and repeatability of an automated robotic ultrasound system (ARTHUR V.2.0) combined with an AI model (DIANA V.2.0) in assessing synovial hypertrophy (SH) and Doppler activity in rheumatoid arthritis (RA) patients, using an expert rheumatologist's assessment as the reference standard.

Methods

30 RA patients underwent two consecutive ARTHUR V.2.0 scans and rheumatologist assessment of 22 hand joints, with the rheumatologist blinded to the automated system's results. Images were scored for SH and Doppler by DIANA V.2.0 using the EULAR-OMERACT scale (0-3). The agreement was evaluated by weighted Cohen's kappa, percent exact agreement (PEA), percent close agreement (PCA) and binary outcomes using Global OMERACT-EULAR Synovitis Scoring (healthy

SUBMITTER: Frederiksen BA 

PROVIDER: S-EPMC12336591 | biostudies-literature | 2025 Aug