A robot arm-assisted acupuncture system with motion and force monitoring: establishment and validation.
Ontology highlight
ABSTRACT: Traditional manual acupuncture faces challenges in quantifying stimulation parameters and preserving techniques consistency. This study presents a robot arm-assisted acupuncture system replicating manual manipulations with precise motion and force monitoring. System validation was conducted through a Laser Vibrometer to measure multiple frequency-amplitude combinations, followed by comparative analysis between robotic and manual acupuncture manipulations using both motion and force monitoring. Additionally, therapeutic efficacy was evaluated through an adjuvant arthritis rat model, comparing the analgesic effects between robotic and manual acupuncture at the Zusanli acupoint. The Laser Vibrometer validation showed high precision, with mean absolute errors below 0.2 mm across all frequency-