Unknown

Dataset Information

Very high-power short-duration using 70W and a flexible tip ablation catheter for pulmonary vein isolation: the POWER PULSE randomized controlled trial.


ABSTRACT:

Aims

Very high-power short-duration (vHPSD) was developed to optimize radiofrequency ablation for atrial fibrillation (AF). However, data on vHPSD ≥ 70W remains limited. We investigated acute efficacy, safety and long-term rhythm outcomes of vHPSD-70W in a randomized controlled trial.

Methods and results

A total of n = 200 patients with paroxysmal AF were randomly assigned 1:1 to receive pulmonary vein isolation (PVI) using vHPSD (70 W/5-7 s) or standard (30-40W, 20-40 s) ablation with a flexible, enhanced-irrigation tip catheter. Primary endpoint was the number of reconnected pulmonary veins (rPV) after adenosine testing. Secondary endpoints included first-pass isolation (FPI), silent cerebral lesions (SCLs) and rhythm outcomes on 12-month follow-up. Mean number of rPVs was

SUBMITTER: Popa MA 

PROVIDER: S-EPMC12365637 | biostudies-literature | 2025 Aug

REPOSITORIES: biostudies-literature