Combined single- and dual-energy CT workflow for dose calculation in radiotherapy.


ABSTRACT:

Background and purpose

Dual-energy computed tomography (DECT) is increasingly used in radiotherapy delineation due to its enhanced soft tissue contrast. DECT also supports direct dose calculation. However, as most current DECT scanners allow for use in only certain body regions, conventional single-energy computed tomography (SECT) is still needed for some patients. A safe clinical introduction of DECT thus requires a combined workflow. This study therefore investigates whether a unified Hounsfield look-up table (HLUT) can be applied across SECT and DECT reconstructions. Patient/material and methods: A Gammex Advanced Electron Density phantom containing tissue-equivalent inserts was scanned using SECT (70-140 kVp and Sn100-Sn140 kVp, Sn meaning tin-filtered) and dual-spiral DECT to identify matching HLUTs for three SECT methods, including a standard reconstruction (only 120 kVp; Method 1), and kVp-independent reconstructions providing mass density (MD; Method 2) or relative electron density (RED; Method 3). Dose agreement was subsequently tested on two anthropomorphic phantoms. For each SECT method, DECT reconstructions were compared through voxel-wise analysis of computed tomography (CT) numbers, and by performing dose calculations in three anatomical regions: head, thorax, and abdomen/pelvis.

Results

Across all three SECT methods, DECT reconstructions with acceptable clinical CT number agreement were identified. Corresponding dose calculations between SECT- and DECT-based plans showed minimal differences.

Interpretation

This phantom study demonstrates that a unified HLUT can be applied across SECT and DECT using standard 120 kVp, MD, or RED reconstructions. This approach may streamline clinical workflows and support a safe and practical transition to DECT-based treatment planning.

SUBMITTER: Sand H 

PROVIDER: S-EPMC12371750 | biostudies-literature | 2025 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Combined single- and dual-energy CT workflow for dose calculation in radiotherapy.

Sand Hella H   Edmund Jens J   Appelt Ane A   Wohlfahrt Patrick P   Taasti Vicki Trier VT   Poulsen Laurids Østergaard LØ   Søndergaard Jimmi J   Nielsen Martin Skovmos MS  

Acta oncologica (Stockholm, Sweden) 20250818


<h4>Background and purpose</h4>Dual-energy computed tomography (DECT) is increasingly used in radiotherapy delineation due to its enhanced soft tissue contrast. DECT also supports direct dose calculation. However, as most current DECT scanners allow for use in only certain body regions, conventional single-energy computed tomography (SECT) is still needed for some patients. A safe clinical introduction of DECT thus requires a combined workflow. This study therefore investigates whether a unified  ...[more]