Feasibility of strong diffusion encoding and fast readout using a plug-and-play head gradient insert at 7 T.


ABSTRACT:

Purpose

Diffusion weighting is achieved by the application of external field gradients typically for tens of milliseconds, during which the signal also substantially decays due to inherent T2 relaxation. Employing strong gradients will shorten the echo time (TE) and increase the signal-to-noise ratio. This study demonstrates a proof-of-principle of using an ultrastrong head gradient insert for diffusion MRI at 7 T.

Methods

A lightweight (45 kg) single-axis gradient coil operating in the z-direction was interfaced with a 7 T system and operated as an additional fourth-gradient axis to the whole-body gradient setup. Several diffusion MRI experiments were conducted to compare acquisitions taken with a gradient insert operating at full capacity (200 mT/m and 1300 T/m/