ABSTRACT: Purpose
A novel 3D-printed phantom design and methodology are proposed, addressing important requirements for technical validation, quality assurance, and multi-site harmonization of quantitative DCE-MRI measurements.Methods
Phantoms were produced by 3D-printing (3DP) gels incorporating channels and pores as proxies for blood vessels and extravascular extracellular space, respectively. A flow circuit was designed to reproduce clinically relevant arterial input functions. Using nine gels with variable porosity and channel size, we evaluated the effect of 3DP parameters on DCE-MRI parameters obtained using the extended Tofts model (ET). Physical gel and fitted model parameters were correlated by multiple linear regression.Results
All phantoms generated realistic arter