Ontology highlight
ABSTRACT: Background
In 2021, Boston Scientific (BSC) announced that its Ingenio pacemakers (PMs) could exhibit high internal battery impedances (HIBIs) and trigger the safety mode (SM) that should preserve pacing when PM function is jeopardized. SM is VVI or biventricular pacing, 72.5 bpm, and unipolar, with a sensitivity of 0.25 mV. In 2023, BSC said SM may pause pacing and 3 deaths occurred in PM-dependent patients. In 2024, BSC warned that some Accolade PMs had HIBI and SM could occur.Objective
This study aimed to quantify adverse events (AEs) associated with SM and compare them with Abbott's backup (BU), Medtronic's reset (RE), and Biotronik's backup mode.Methods
The Food and Drug Administration's Manufacturer and User Facility Device Experience database was searched fro
SUBMITTER: Hauser RG
PROVIDER: S-EPMC12411951 | biostudies-literature | 2025 Aug
REPOSITORIES: biostudies-literature