ABSTRACT: Purpose
To investigate whether performing phacoemulsification with a lower infusion pressure using the Centurion active sentry system affects surgical efficiency, complications and a range of clinical and imaging parameters compared to the higher pressures routinely used in patients with cataract and concomitant diabetic retinopathy and glaucoma.Setting
Sunderland Eye Infirmary, Sunderland, United Kingdom.Design
Masked observer randomized controlled feasibility trial.Methods
Patients with cataracts undergoing routine phacoemulsification with either diabetic retinopathy or primary open-angle glaucoma of any severity were included and randomized to an infusion pressure of 30 ('LOW') or 60 ('HIGH') mmHg. All other fluidic settings were standardized. Surgical me