Ontology highlight
ABSTRACT: Background
Magnetic resonance imaging (MRI) is susceptible to motion artifacts, a particular challenge in pediatric imaging. External motion tracking devices and navigator techniques have been previously explored, but challenges persist necessitating sedation.Purpose
To establish a new awake brain-MRI exam tolerated by children with diagnostic quality images.Materials and methods
Participants were prospectively recruited and investigated using an in-house developed multi-sequence scan called NeuroMix that produces T1-weighted, T2-weighted, T2*-weighted, T2-FLAIR, and diffusion-weighted images in under 3 minutes. Additionally, a self-created motion tracking device was attached to participants' foreheads to perform prospective motion correction (PMC) on 2D single-shot
SUBMITTER: van Niekerk A
PROVIDER: S-EPMC12429249 | biostudies-literature | 2024 May