Quantifying lower extremity blood flow using low-dose CT perfusion: validation in a swine model.


ABSTRACT:

Background

Quantitative assessment of blood flow in peripheral extremities in conjunction with simultaneous CT angiography measurements can improve risk assessment and provide a critical decision-making tool for patients across a wide spectrum of vascular disease severity.

Purpose

This study assessed the reproducibility and accuracy of lower extremity blood flow measurements with a low-dose first-pass analysis CT perfusion technique.

Materials and methods

This prospective study utilized 16 Yorkshire Swine to obtain lower extremity blood flow CT measurements at baseline and under induced femoral stenosis using a vascular occluder. Thirty-three pairs of CT measurements evaluated reproducibility, and 43 CT measurements assessed accuracy against ultrasound flow probe refe