ABSTRACT: Objective
A significant anatomical limitation of transcatheter mitral valve replacement (TMVR) is the risk of left ventricular outflow tract obstruction. Here, we describe balloon-assisted translocation in mitral anterior leaflet (BATMAN) in a patient with a calcified bioprosthetic mitral valve.Key steps
The anterior mitral leaflet is traversed with a 0.014-inch stiff wire with electrocautery, followed by the delivery of a long sheath and placement of dual wires. The dual-wire system allows for BATMAN, followed by a rapid delivery of the transcatheter heart valve, minimizing exposure to significant mitral regurgitation.Potential pitfalls
Leaflet traversal with a 0.035-inch system may be difficult because of leaflet calcification, and translocation of the anterior mi