Combining Lumipulse p-tau217 and Aβ42/40 as confirmatory tests for Aβ positivity prior to disease-modifying therapy.


ABSTRACT:

Introduction

For a blood-based biomarker to be considered a confirmatory test for the detection of abnormal amyloid beta (Aβ) levels, the sensitivity and specificity must be equivalent to that of current cerebrospinal fluid tests.

Methods

In the current study we assessed the ability of phosphorylated tau (p-tau)217 and Aβ42/40 from the Lumipulse G p-tau217 and β-amyloid ratio (1-42/1-40) tests, individually and combined, to predict Aβ positron emission tomography status in two sub-cohorts from the Australian Imaging, Biomarkers, and Lifestyle Study of Ageing.

Results

Testing an Alzheimer's disease continuum cohort, the area under the curve (AUC), sensitivity, specificity, and accuracy for the p-tau217/Aβ42 ratio reached 0.961, 93%, 92%, and 93%, respectively. Validati

SUBMITTER: Doecke JD 

PROVIDER: S-EPMC12450607 | biostudies-literature | 2025 Sep

REPOSITORIES: biostudies-literature