The Safety and Performance of a Novel Extracorporeal Membrane Oxygenation Device in a Long-Term Ovine Model.
Ontology highlight
ABSTRACT: Since extracorporeal membrane oxygenation (ECMO) is primarily used for patients in a high-risk state and is an invasive procedure, its unique application scenarios make it difficult to recruit suitable cases for clinical trials. Therefore, large animal models have become one of the most important models for preclinical evaluation of the safety and effectiveness of ECMO. This study aims to assess the safety and performance of a novel portable ECMO device with Small-tail Han sheep. Fifteen sheep were divided into a test group (LIFEMOTION, Chinabridge, Shenzhen, China) and control group (NOVALUNG XLUNG kit 230, Xonis, Heilbronn, Germany) with veno-venous ECMO (VV-ECMO) and veno-arterial ECMO (VA-ECMO) modes. Tracheal intubation, arteriovenous access, and ECMO support were performed. Vital sig
SUBMITTER: Li Y
PROVIDER: S-EPMC12452530 | biostudies-literature | 2025 Sep
REPOSITORIES: biostudies-literature