Measurement accuracy of CT systems: The importance of calibration phantoms.
Ontology highlight
ABSTRACT: This study aims to evaluate the measurement accuracy of computed tomography (CT) systems, focusing on the necessity of using calibration phantoms for enhanced precision. Both clinical CT and micro-CT systems were evaluated using a specially designed two-ball phantom, which provides a reliable reference for spatial resolution and geometric accuracy. The study involved scanning the phantom with two micro-CT devices (the oversize micro-CT SkyScan 1173 and the high-resolution micro-CT SkyScan 1272) and a clinical CT device, a third-generation dual-source CT scanner (SOMATOM Force), measuring the distance between the centres of two ruby balls. The results showed significant differences in measurement accuracy between the devices. The high-resolution micro-CT provided the most consistent measure
SUBMITTER: Scherberich J
PROVIDER: S-EPMC12463237 | biostudies-literature | 2025
REPOSITORIES: biostudies-literature
ACCESS DATA