Reliability of an all-in-one wearable sensor for continuous vital signs monitoring in high-risk patients: the NIGHTINGALE clinical validation study.
Ontology highlight
ABSTRACT: Continuous vital signs monitoring with wearable systems may improve early recognition of patient deterioration on hospital wards. The objective of this study was to determine whether the wearable Checkpoint Cardio's CPC12S, can accurately measure heart rate (HR), respiratory rate (RR), oxygen saturation (SpO2), blood pressure (BP) and temperature continuously. In an observational multicenter method comparison study of 70 high-risk surgical patients admitted to high-dependency wards; HR, RR, SpO2, BP and temperature were simultaneously measured with the CPC12S system and with ICU-grade monitoring systems in four European hospitals. Outcome measures were bias and 95% limits of agreement (LoA). Clinical accuracy was assessed with Clarke Error Grid analyses for HR and RR. A total of 3,212 h of
SUBMITTER: Breteler MJM
PROVIDER: S-EPMC12474673 | biostudies-literature | 2025 Oct
REPOSITORIES: biostudies-literature